FDAClass IIFebruary 1, 2017

Tina-Quant recall

Tina-Quant (Latex) HS Test System (C-Reactive Protein (Latex) High Sensitive) The Tina-Quant CRP (Latex) High Sensitive Immunoturbidimetric assay is…

Why was Tina-Quant recalled?

Tina-quant Cardiac C-reactive Protein (Latex) High Sensitive (CRPHS) lot number 604450 (catalog number 11972855216) shows an under-recovery of up to -25% at concentrations below 5 mg/L on the MODULAR ANALYTICS P module. this issue can lead to erroneous low CRPHS results. CRPHS is used as a marker for the prediction of cardiovascular risk. An erroneous low result could lead to an incorrect cardiovascular risk assessment and subsequent delay of further examination.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog number: 11972855216, lot number 60445001

Product
Tina-Quant (Latex) HS Test System (C-Reactive Protein (Latex) High Sensitive) The Tina-Quant CRP (Latex) High Sensitive Immunoturbidimetric assay is…
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
3204
Distribution
Nationwide Distribution including AL, AZ, CA, CT, FL, GA, KS, MA, MO, NJ, NY, PA, and WA.
Recall date
February 1, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls