00434811113 recall
00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm Stem Diameter 130 mm Stem Length Trabecular Metal Humeral…
Why was 00434811113 recalled?
This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength approximately 80% of the minimum required by specification in one of the three zones of the Trabecular Metal. The other two zones exceeded the minimum specification.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
lot 62267250
- Product
- 00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm Stem Diameter 130 mm Stem Length Trabecular Metal Humeral…
- Recalling firm
- Zimmer, Inc.
- Quantity
- 10 units
- Distribution
- Distribution USA in the state of Illinois.
- Recall date
- March 5, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)