FDAClass IIMarch 5, 2014

The FOCUS is a microprocessor controlled intra-oral recall

The FOCUS is a microprocessor controlled intra-oral x-ray unit with a Load Matching Frequency (LMF) DC generator, which produces dental images on…

Why was The FOCUS is a microprocessor controlled intra-oral recalled?

Instrumentarium Dental has delivered a number of FOCUS 3.0 Intraoral X-ray units in November 2013 with a potential flaw in the mounting hardware. A number of wall mount units were delivered with FOCUS X-ray machines that had not been properly welded at our supplier. As a result of the incomplete welding operation, the wall mount unit on the FOCUS X-ray unit may not properly support the unit as expected, which could cause the unit to drift. Depending on the extent of the shortened or missing welds, there is also a possibility that the wall mount unit will fail to support the weight of the unit, resulting in the FOCUS X-ray unit falling unexpectedly from its mounted position.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Device Model Number / Sales Code: 50600-IMG; Serial Numbers (US and Foreign): F21221 - F21641, and F21643; Serial Numbers (US Devices): F21221 - F21230, F21265 - F21270, F21273 - F21290, F21296 - F21350, F23154 - F21365, F21367 - F21391, F21395 - F21447, F21451 - F21464, F21469 - F21478, F21481 - F21486, F21490 - F21519, F21525 - F21542, F21545 - F21550, F21565 - F21588, F21623 - F21628, F21636 - F21641, and F21643

Product
The FOCUS is a microprocessor controlled intra-oral x-ray unit with a Load Matching Frequency (LMF) DC generator, which produces dental images on…
Recalling firm
Instrumentarium Dental, PaloDEx Group Oy
Quantity
101 units (US Distribution)
Distribution
Worldwide Distribution: US (Nationwide) including states of: AK, AZ, CA, CO, GA, IL, KY, MD, MT, NC, NV, OR, PA, TN, TX, UT, VA, WA,and WI; and Internationally to: Belgium Brazil, Chile, China, Czech Republic, France, Germany, Italy, Spain, Sweden, Switzerland, and United Kingdom.
Recall date
March 5, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls