FDAClass IIFebruary 21, 2024

CombiDiagnost R90 1.1 recall

CombiDiagnost R90 1.1 (709031)

Why was CombiDiagnost R90 1.1 recalled?

While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model: R90 1.1 (709031); UDI DI: 00884838101456; Software version 1.1.1 or higher; Serial Numbers: 10001004, 10001005, 10001006, 10001008, 10001009, 10001010, 10001011, 10001012, 10001013, 10001018, 10001020, 10001022, 10001024, 10001026, 10001027, 10001028, 10001029, 10001030, 10001031, 10001032, 10001035, 10001036, 10001043, 10001044, 10001045, 10001046, 10001047, 10001048, 10001049, 10001050, 10001051, 10001052, 10001053, 10001054, 10001055, 10001056, 10001057, 10001058, 10001059, 10001060, 10001061, 10001062, 10001063, 10001064, 10001065, 10001066, 10001067, 10001068, 10001069, 10001070, 10001071, 10001072, 10001073, 10001074, 10001075, 10001076, 10001077, 10001078, 10001079, 10001080, 10001081, 10001083, 10001084, 10001085, 10001086, 10001087, 10001088, 10001089, 10001090, 10001091, 10001092, 10001093, 10001094, 10001095, 10001096, 10001097, 10001098, 10001099, 10001100, 10001101, 10001102, 10001103, 10001104, 10001105, 10001106, 10001107, 10001108, 10001109, 10001110, 10001111, 10001112, 10001113, 10001114, 10001115, 10001116, 10001117, 10001118, 10001119, 10001120, 10001121, 10001122, 10001123, 10001124, 10001125, 10001126, 10001127, 10001128, 10001129, 10001130, 10001131, 10001132, 10001133, 10001134, 10001135, 10001136, 10001137, 10001138, 10001139, 10001140, 10001141, 10001142, 10001143, 10001144, 10001145, 10001147, 10001148, 10001149, 10001150, 10001152, 10001201;

Product
CombiDiagnost R90 1.1 (709031)
Recalling firm
PHILIPS MEDICAL SYSTEMS
Quantity
130 units
Distribution
US Nationwide distribution in the states of CA, FL, GA, IL, IN, KS, MD, ME, MI, MN, NC, NJ, NY, TX, VA, WA.
Recall date
February 21, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls