MOSAIQ Oncology Information System with recall
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Why was MOSAIQ Oncology Information System with recalled?
Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
MOSAIQ v 3.10.200, and 3.1.1.0 up to and including v 3.2.3.1 with Particle Base. Software Versions: 3.10.200, 3.1.1.0, 3.1.2.0, 3.1.3.0, 3.1.3.1, 3.1.3.2, 3.1.3.3, 3.1.3.4, 3.2.0.0, 3.2.1.0, 3.2.1.1, 3.2.1.2, 3.2.1.3, 3.2.1.4, 3.2.2.0, 3.2.2.1, 3.2.3.0, 3.2.3.1 UDI-DI: 07340201500026, 07340201500071
- Product
- MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
- Recalling firm
- Elekta, Inc.
- Quantity
- 1
- Distribution
- US Nationwide distribution in the state of TX.
- Recall date
- January 21, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)