FDAClass IIJanuary 21, 2026

MOSAIQ Oncology Information System with recall

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Why was MOSAIQ Oncology Information System with recalled?

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

MOSAIQ v 3.10.200, and 3.1.1.0 up to and including v 3.2.3.1 with Particle Base. Software Versions: 3.10.200, 3.1.1.0, 3.1.2.0, 3.1.3.0, 3.1.3.1, 3.1.3.2, 3.1.3.3, 3.1.3.4, 3.2.0.0, 3.2.1.0, 3.2.1.1, 3.2.1.2, 3.2.1.3, 3.2.1.4, 3.2.2.0, 3.2.2.1, 3.2.3.0, 3.2.3.1 UDI-DI: 07340201500026, 07340201500071

Product
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Recalling firm
Elekta, Inc.
Quantity
1
Distribution
US Nationwide distribution in the state of TX.
Recall date
January 21, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls