Getinge 9100-Series Washer Disinfector recall
Getinge 9100-Series Washer Disinfector models: 9120E, 9125E, 9127E, 9128E, 9122EW, 9128EW.
Why was Getinge 9100-Series Washer Disinfector recalled?
Inadequate documentation verifying whether device installation has been executed in accordance with the manufacturer's requirements. This can lead to installation failure modes and hazardous situations.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI Codes: [07340153710023, 07340153710047, 07340153710054, 07340153710061, 9122EW 07340153710030, 07340153710078.] Serial Numbers: WAA088008, WAA088276, WAA089054, WAA089188, WAA089511, WAA091312, WAA092134, WAA097871, WAA098978, WAA087395, WAA087675, WAA087676, WAA088563, WAA088677, WAA088684, WAA088798, WAA088973, WAA090326, WAA090827, WAA090829, WAA091562, WAA092656, WAA092905, WAA093809, WAA094122, WAA094447, WAA095653, WAA095654, WAA095655, WAA096084, WAA096723, WAA096820, WAA099698, WAA100249, WAA100596, WAA100828, WAA102416, WAA102300, WAA102301, WAA086373, WAA088830, WAA089854, WAA089735, WAA089736, WAA090141, WAA090273, WAA090274, WAA091427, WAA093786, WAA094342, WAA095013, WAA095014, WAA096724, WAA098589, WAA099164, WAA098725, WAA100514, WAA100695, WAA100704, WAA100832, WAA101375, WAA100834, WAA101893, WAA076270, WAA085559, WAA086376, WAA086914, WAA086915, WAA088007, WAA088273, WAA088464, WAA088465, WAA087754, WAA088678, WAA088685, WAA088871, WAA088998, WAA089258, WAA089356, WAA089446, WAA089970, WAA089969, WAA090495, WAA090615, WAA090746, WAA091561, WAA091184, WAA091560, WAA091563, WAA091564, WAA091869, WAA091870, WAA091873, WAA091868, WAA092136, WAA091872, WAA092904, WAA093646, WAA093882, WAA093967, WAA094123, WAA094202, WAA094284, WAA094285, WAA094286, WAA094617, WAA094601, WAA095421, WAA096425, WAA096427, WAA096428, WAA097298, WAA098056, WAA099694, WAA099450, WAA099451, WAA099695, WAA099697, WAA100127, WAA100826, WAA100827, WAA100830, WAA101342, WAA101894, WAA103173, WAA088274, WAA093968
- Product
- Getinge 9100-Series Washer Disinfector models: 9120E, 9125E, 9127E, 9128E, 9122EW, 9128EW.
- Recalling firm
- Getinge Usa Sales Inc
- Quantity
- 127
- Distribution
- Worldwide distribution: (US) AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, SD, TX, VA, WA, WV. (OUS) Australia, Austria, Belgium, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Italy, Japan, Kuwait, Libya, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
- Recall date
- May 25, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)