FDAClass IIFebruary 1, 2017

Sekisui Diagnostics Ammonia L3K¿ Assay recall

Sekisui Diagnostics Ammonia L3K¿ Assay; Catalog Number: 293-80-91 and 293-10 For the IN VITRO quantitative measurement of ammonia concentration in…

Why was Sekisui Diagnostics Ammonia L3K¿ Assay recalled?

Certain lots are showing an atypical decrease in optical density (OD) over time, which may result in an impact to performance at the high end of the linear range.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog Number: 293-80-91; Lot Numbers: 48737, 48962, 49087, 49220, 49258 and Catalog Numbers: 293-10; Lot Numbers: 48967, 49330, 49331, 49515

Product
Sekisui Diagnostics Ammonia L3K¿ Assay; Catalog Number: 293-80-91 and 293-10 For the IN VITRO quantitative measurement of ammonia concentration in…
Recalling firm
Sekisui Diagnostics P.E.I. Inc.
Quantity
774 Liters bulk and 565 kits
Distribution
Worldwide Distribution - US including DE, KS, FL, TX, CA, MI, TN, MA, ID, NJ, RI, IL and Internationally to Austria, Mexico, Canada, India, South Korea, China, Spain, South Korea, and Sri Lanka
Recall date
February 1, 2017
Classification
Class II
Status
Terminated