Allura Xper FD20/20 recall
Allura Xper FD20/20; Model Number: 722038;
Why was Allura Xper FD20/20 recalled?
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Number: 722038; System Product Name: Allura Xper FD20/20; UDI-DI: 00884838054226; Serial Numbers: 11, 117, 110, 134, 64, 90, 124, 72;
- Product
- Allura Xper FD20/20; Model Number: 722038;
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Quantity
- 8 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.
- Recall date
- January 21, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)