FDAClass IIJanuary 21, 2026

Allura Xper FD10 recall

Allura Xper FD10; Model Number: 722010;

Why was Allura Xper FD10 recalled?

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Number: 722010; System Product Name: Allura Xper FD10; UDI-DI: 00884838059030; Serial Numbers: 764, 1396, 728, 980, 782, 125, 1096, 661, 225, 1055, 1185, 368, 743, 1006, 570, 901, 552, 973, 795, 888, 1140, 781, 1060, 354, 186, 516, 517, 64, 386, 387, 530, 996, 964, 1265, 525, 307, 1189, 98, 774, 935, 721, 254, 1008, 1213, 1273, 66, 156, 304, 1422, 388, 235, 236, 1066, 1090, 549, 689;

Product
Allura Xper FD10; Model Number: 722010;
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Quantity
56 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.
Recall date
January 21, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)