FDAClass IIFebruary 1, 2017

Apollo DRF recall: Labeling Error

Apollo DRF: Model: 9784220831 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.

Why was Apollo DRF recalled?

The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

08090531 09090591 10040680 15051450 11020798

Product
Apollo DRF: Model: 9784220831 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
Recalling firm
Villa Radiology Systems LLC
Quantity
66 distributed in US
Distribution
AL AZ CA CO CT DC FL IA IL IN LA MA MD MI MN MO MS NJ NY OH PA Puerto Rico TN TX UT WI
Recall date
February 1, 2017
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls