FDAClass IIMay 18, 2022

11272VN 11272VNK Flexible Video Urethro Cystoscope recall

11272VN 11272VNK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018); 11272VNU 11272VNUK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018)

Why was 11272VN 11272VNK Flexible Video Urethro Cystoscope recalled?

Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

11272VN 11272VNK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018) UDI Code: 4048551226186; 11272VNU 11272VNUK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018) UDI Code: 4048551226193; All Serial Numbers manufactured/distributed since January 2018.

Product
11272VN 11272VNK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018); 11272VNU 11272VNUK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018)
Recalling firm
Karl Storz Endoscopy
Quantity
All Serial Numbers manufactured/distributed since January 2018
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA,WI, WV, and WY. O.U.S.: Mexico, Lebanon, and South Korea
Recall date
May 18, 2022
Classification
Class II
Status
Ongoing
Agency
FDA (device)