Apollo recall: Labeling Error
Apollo: Models: 9784000131, 9784000231, 9784020131, 9784020231, 9784100231, 9784120131, 9784120231, 9784130131 Multi-function x-ray systems capable of…
Why was Apollo recalled?
The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
06040078 07030172 06090116 06110138 07010149 07020158 07020162 07020163 07020165 07030168 07030171 07040206 09090605 10100752 06060086 06100127 09010566 08020473 11100873 15041446
- Product
- Apollo: Models: 9784000131, 9784000231, 9784020131, 9784020231, 9784100231, 9784120131, 9784120231, 9784130131 Multi-function x-ray systems capable of…
- Recalling firm
- Villa Radiology Systems LLC
- Quantity
- 66 distributed in US
- Distribution
- AL AZ CA CO CT DC FL IA IL IN LA MA MD MI MN MO MS NJ NY OH PA Puerto Rico TN TX UT WI
- Recall date
- February 1, 2017
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)