Persona recall
Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04. Product Usage: The…
Why was Persona recalled?
Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this, TASP shims must be manually cleaned and inspected per product insert, 87-6203-991-22.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04: 42-5279-002-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 10mm; 42-5279-002-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 12mm; 42-5279-002-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 14mm; 42-5279-004-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 10mm; 42-5279-004-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 12mm; 42-5279-004-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 14mm; 42-5279-006-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 10mm; 42-5279-006-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 12mm; 42-5279-006-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 14mm; 42-5279-008-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 10mm; 42-5279-008-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 12mm; 42-5279-008-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 14mm; 42-5279-003-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 10mm; 42-5279-003-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 11mm; 42-5279-003-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 12mm; 42-5279-003-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 13mm; 42-5279-003-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 14mm; 42-5279-005-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 10mm; 42-5279-005-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 11mm; 42-5279-005-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 12mm; 42-5279-005-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 13mm; 42-5279-005-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 14mm; 42-5279-007-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 10mm; 42-5279-007-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 11mm; 42-5279-007-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 12mm; 42-5279-007-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 13mm; 42-5279-007-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 14mm.
- Product
- Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04. Product Usage: The…
- Recalling firm
- Zimmer, Inc.
- Quantity
- 67963
- Distribution
- Worldwide Distribution - USA Nationwide including DC, and PR and the countries of Australia, Austria, Belgium, Canada, Czech Republic, Dominican Republic, France, Germany, India, Iran, Israel, Italy, Japan, Luxembourg, Netherlands, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
- Recall date
- February 11, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)