FDAClass IIMay 18, 2022

Philips Hemodynamic Application recall

Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient…

Why was Philips Hemodynamic Application recalled?

When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI: B-00884838084544; System Serial Numbers: 396 411 225 351 51 46 33 189 132 19 51 924 644 645 646 376 384 487 538 886 14 15 643 1194 220 1238 947 949 950 951 89 657 169 168 413 414 624 565 1060 393 512 278 818 362 123 802 395 576 932 291 834 539 605 161 970 686 761 229 230

Product
Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient…
Recalling firm
Philips North America Llc
Quantity
77
Distribution
Devices were distributed to the following US states: FL, ID, NJ, OK, SC and TX. Devices were distributed to the following foreign countries: France, Germany, India, Mexico, Netherlands, New Zealand, Norway, Poland, South Africa, and Switzerland.
Recall date
May 18, 2022
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls