CR Precoat FEM Comp recall
CR PRECOAT FEM COMP SIZE BMR CR PRECOAT FEM COMP SIZE C/R CR PRECOAT FEM COMP SIZE D/R CR PRECOAT FEM COMP SIZE E/R CR PRECOAT FEM COMP SIZE F/R CR…
Why was CR Precoat FEM Comp recalled?
The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
00-5972-015-01 61777593 00-5972-018-02 4002192 00-5970-012-02 61878272 00-5970-013-02 61777576 00-5970-013-02 62135909 00-5970-014-02 11400192 00-5970-014-02 62122787 00-5970-014-02 62135920 00-5970-014-02 62132399 00-5970-014-02 62132399R 00-5970-014-02 62135921 00-5970-015-02 11002229 00-5970-015-02 61777581 00-5970-015-02 62135936 00-5970-016-02 61777583 00-5970-016-02 62128113 00-5970-017-01 11003049 00-5970-018-01 61758971 00-5970-018-01 11003323 00-5970-018-02 11003330 00-5970-018-02 11003322
- Product
- CR PRECOAT FEM COMP SIZE BMR CR PRECOAT FEM COMP SIZE C/R CR PRECOAT FEM COMP SIZE D/R CR PRECOAT FEM COMP SIZE E/R CR PRECOAT FEM COMP SIZE F/R CR…
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 13,227 in total
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.
- Recall date
- March 21, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)