Cobas b 221 recall: Contamination
Cobas b 221 (versions 5 and 6) Blood Gas and Electrolytes Analyzer
Why was Cobas b 221 recalled?
Chemical component of the F3 Fluid pack and environmental bacterial contamination causes a positive bias up to 39% at 4.1 mmol/l for the glucose measurement and 30% at 4.12 mmol/l for lactate. Urea readings are unaffected.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model/Catalog/Part #: 1. cobas b221 <6>=Roche OMNI S6 SYSTEM; and 2. 03337154692 - cobas b 221<6>=Roche OMNI S6 SYSTEM CU Potassium - CEM, Calcium - JFP, Sodium - JGS, Chloride - CGZ, Blood Gases (PCO2, PO2) and pH - CHL, Hemoglobin -GKR, Hematocrit - GKF, Oxygen saturation, Carboxyhemoglobin - GHS, Oxyhemoglobin - GGZ, Methemoglobin - KHG, Deoxyhemoglobin - GKA, Lactic Acid - KHP, Glucose- CGA, Urea Nitrogen - CDS.
- Product
- Cobas b 221 (versions 5 and 6) Blood Gas and Electrolytes Analyzer
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- 103
- Distribution
- US Nationwide distribution.
- Recall date
- February 11, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)