FDAClass IIFebruary 11, 2015

Cobas b 221 recall: Contamination

Cobas b 221 (versions 5 and 6) Blood Gas and Electrolytes Analyzer

Why was Cobas b 221 recalled?

Chemical component of the F3 Fluid pack and environmental bacterial contamination causes a positive bias up to 39% at 4.1 mmol/l for the glucose measurement and 30% at 4.12 mmol/l for lactate. Urea readings are unaffected.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model/Catalog/Part #: 1. cobas b221 <6>=Roche OMNI S6 SYSTEM; and 2. 03337154692 - cobas b 221<6>=Roche OMNI S6 SYSTEM CU Potassium - CEM, Calcium - JFP, Sodium - JGS, Chloride - CGZ, Blood Gases (PCO2, PO2) and pH - CHL, Hemoglobin -GKR, Hematocrit - GKF, Oxygen saturation, Carboxyhemoglobin - GHS, Oxyhemoglobin - GGZ, Methemoglobin - KHG, Deoxyhemoglobin - GKA, Lactic Acid - KHP, Glucose- CGA, Urea Nitrogen - CDS.

Product
Cobas b 221 (versions 5 and 6) Blood Gas and Electrolytes Analyzer
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
103
Distribution
US Nationwide distribution.
Recall date
February 11, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls