2.7 X 6 Cort Screw/selfta 2.7 X 7 Cort Screw/selfta 2.7 recall: Sterility Concern
2.7 X 6 CORT SCREW/SELFTA 2.7 X 7 CORT SCREW/SELFTA 2.7 X 8 CORT SCREW/SELFTA 2.7 X 9 CORT SCREW/SELFTA 2.7 X 10 CORT SCREW/SELFT 2.7 X 12 CORT…
Why was 2.7 X 6 Cort Screw/selfta 2.7 X 7 Cort Screw/selfta 2.7 recalled?
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
61173335 61173337 61173339 61173340 61173341 61115956 61173344 61173346 61173347 61173349 61173350 61173352 61173353 61173354 61173355 61173356 61173357 61173358 61173359 61173360 61173361 61173362 61173363 61173364 61173365 61173366 61359816
- Product
- 2.7 X 6 CORT SCREW/SELFTA 2.7 X 7 CORT SCREW/SELFTA 2.7 X 8 CORT SCREW/SELFTA 2.7 X 9 CORT SCREW/SELFTA 2.7 X 10 CORT SCREW/SELFT 2.7 X 12 CORT…
- Recalling firm
- Zimmer, Inc.
- Quantity
- 7,044,680 total
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
- Recall date
- February 26, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)