FDAClass IIFebruary 8, 2023

Non-sterile procedural trays labeled as MAJOR/MINOR recall: Sterility Concern

Non-sterile procedural trays labeled as MAJOR/MINOR, 3 per case

Why was Non-sterile procedural trays labeled as MAJOR/MINOR recalled?

Potential for sterile blades, within non-sterile SPT kits containing the blades, to puncture the outer foil layer causing a breach in the sterile packaging.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Item Number: DYNJ900416M, UDI/GTIN Case: 40195327031672, UDI/GTIN Each: 10195327031671, Lot Number: 22IBQ980

Product
Non-sterile procedural trays labeled as MAJOR/MINOR, 3 per case
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
160 cases (480 units)
Distribution
US Nationwide distribution in the states of AR, MD, WA, WI.
Recall date
February 8, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls