FDAClass IIApril 10, 2013

Integra Ojemann Cortical Stimulator recall

Integra Ojemann Cortical Stimulator; Product / Catalogue No: OCS2. Intended for intraoperative cortical stimulation mapping procedures.

Why was Integra Ojemann Cortical Stimulator recalled?

If the headphone jack is in use during a surgical procedure with the OCS2 and a non-intended voltage (such as static electricity) comes in contact with the outer case of the unit, that voltage could be transmitted to the patient and could be a potential source of injury.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Number Range: 1010 through 1380

Product
Integra Ojemann Cortical Stimulator; Product / Catalogue No: OCS2. Intended for intraoperative cortical stimulation mapping procedures.
Recalling firm
Integra Burlington MA, Inc.
Quantity
371 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Austria, Belgium, Czech Republic, Belgium, Denmark, Spain, Great Britain, Hungary, Ireland, Italy, Netherlands, Poland, Morocco, Netherlands, Poland, Portugal, Saudia Arabia, Sweden, Turkey, and South Africa.
Recall date
April 10, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls