FDAClass IIFebruary 4, 2015

Syngo.plaza: Syngo.plaza is a Picture Archiving and recall

Syngo.plaza: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute…

Why was Syngo.plaza: Syngo.plaza is a Picture Archiving and recalled?

In case of a system crash, images may not be written to the hard disk and this may result in inconsistencies in the database. In case of a system crash (e.g. blue screen, power outage) images may not be written from cache to the hard disk and might get lost.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Numbers: 10592457, 10863171, 10863173, 10863172; with serial numbers: Serial 100544 100262 100198 100258 200211 100135 100133 100213 100031 100084 100233 100140 200308 100245 100314 100139 100177 100197 100170 100025 100174 100186 100033 100306 100232 100185 100256 100199 100188 100257 100198 100226 100198 100228 100181 100029 100253 100301 100203 100132 100139 100247 100215 100230 100221 100146 100207 100212 100539 100206

Product
Syngo.plaza: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute…
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
50
Distribution
Nationwide Distribution
Recall date
February 4, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)