FDAClass IIJanuary 14, 2026

Beckman Coulter UniCel Dxl 800 recall

Beckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit

Why was Beckman Coulter UniCel Dxl 800 recalled?

Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

973100: UDI 15099590265373 (Chaska) and 16954701000412 (Suzhou); A71456: UDI 15099590369194; A71457: UDI 15099590369231; A25288: UDI 15099590194819

Product
Beckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit
Recalling firm
Beckman Coulter, Inc.
Quantity
6457 units
Distribution
Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Cabo Verde, Canada, Chile, China, Costa Rica, Croatia, Czechia, Denmark, Dominican Republic, Egypt, El Salvador, Eswatini, Ethiopia, France, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lebanon, Libya, Lithuania, Malaysia, Mayotte, Mexico, Moldova, Republic of, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam, and Zimbabwe.
Recall date
January 14, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls