Beckman Coulter UniCel Dxl 800 recall
Beckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit
Why was Beckman Coulter UniCel Dxl 800 recalled?
Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
973100: UDI 15099590265373 (Chaska) and 16954701000412 (Suzhou); A71456: UDI 15099590369194; A71457: UDI 15099590369231; A25288: UDI 15099590194819
- Product
- Beckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 6457 units
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Cabo Verde, Canada, Chile, China, Costa Rica, Croatia, Czechia, Denmark, Dominican Republic, Egypt, El Salvador, Eswatini, Ethiopia, France, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lebanon, Libya, Lithuania, Malaysia, Mayotte, Mexico, Moldova, Republic of, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam, and Zimbabwe.
- Recall date
- January 14, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)