FDAClass IIJanuary 25, 2017

Fresenius 2008T Series Hemodialysis System recall

Fresenius 2008T Series Hemodialysis System

Why was Fresenius 2008T Series Hemodialysis System recalled?

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All serial numbers

Product
Fresenius 2008T Series Hemodialysis System
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Quantity
65361 units (US) 249 units (OUS)
Distribution
Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam
Recall date
January 25, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls