Versys 6 Inch Beaded FC 13x160mm LM Body EXT Neck & STD recall
VERSYS 6 INCH BEADED FC 13X160MM LM BODY EXT NECK & STD NECK Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Why was Versys 6 Inch Beaded FC 13x160mm LM Body EXT Neck & STD recalled?
The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Item Number: 00-7843-013-56 Lot Numbers: 61993635 61987936 62011961 Item Number: 00-7843-013-06 Lot Numbers: 61721714 62097225
- Product
- VERSYS 6 INCH BEADED FC 13X160MM LM BODY EXT NECK & STD NECK Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 13,227 in total
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.
- Recall date
- March 21, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)