FDAClass IIMay 11, 2022

BD Synapsys Microbiology Informatics Solution recall

BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10

Why was BD Synapsys Microbiology Informatics Solution recalled?

When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All systems running BD Synapsys version 4.10; UDI: 00382904441500; Serial Numbers: SL00530 SL00515 SL00527 SL00502 SL00540 SL00564 SL00549 SL00509 SL00003 SL00004 SL00010 SL00001 SL00005 SL00002 SL000015 SL01010 SL01022

Product
BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10
Recalling firm
Becton Dickinson & Co.
Quantity
15 systems
Distribution
Worldwide distribution US Nationwide distribution in the states of CA, IL, MI, NY, OH, SC, and TN. The countries of Austria, France, Germany, Netherlands, Sweden.
Recall date
May 11, 2022
Classification
Class II
Status
Terminated
Agency
FDA (device)