Covidien Trellis-6 Peripheral Infusion System recall
Covidien Trellis-6 Peripheral Infusion System. Models BVT608010V01, BVT608030V01, BVT612010V01, and BVT612030V01. Sterile EO. Product Usage: Intended…
Why was Covidien Trellis-6 Peripheral Infusion System recalled?
A manufacturing error resulted in the risk of incorrect proximal and distal balloon inflation port identification on the units. Units have been identified to have the distal balloon inflation port incorrectly labeled as proximal, and, the proximal balloon port incorrectly labeled as distal.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model BVT608010, Lot numbers: 9853174, 9864497, 9890772, 9890931, 9925676, 9932672, 9937308, 9937315, 9940474, 9940614, 9976340, 9978233, A001589, A001740. Model BVT608030, Lot Numbers: 9853331, 9887597, 9887695, 9887757, 9932042. Model BVT612010, Lot numbers: 9854121, 9868214, 9868311, 9875472, 9875632, 9876174, 9883295, 9884788, 9884951, 9886196, 9886312, 9932096, 9968511, 9974109, 9974126, 9976472, 9977091, 9978862, A002299, A002423. Model BVT312030, Lot numbers: 9854124, 9892081, 9941213, 9971096, 9971139, 9976506, 9977140, 9978964, 9979548, A003021, A003121, A003651, A007900, A008492, A008603, A008687.
- Product
- Covidien Trellis-6 Peripheral Infusion System. Models BVT608010V01, BVT608030V01, BVT612010V01, and BVT612030V01. Sterile EO. Product Usage: Intended…
- Recalling firm
- Covidien
- Quantity
- 216 devices (207 US, 9 OUS)
- Distribution
- Worldwide Distribution - US Nationwide and the countries Australia, Canada, Finland, France, Germany, Ireland, Italy, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
- Recall date
- February 11, 2015
- Classification
- Class I
- Status
- Terminated
- Agency
- FDA (device)