FDAClass IIFebruary 1, 2023

Alinity m System. The Alinity m System is designed to recall

Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification…

Why was Alinity m System. The Alinity m System is designed to recalled?

The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Number: 08N53-002

Product
Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification…
Recalling firm
Abbott Molecular, Inc.
Quantity
918 units
Distribution
x
Recall date
February 1, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls