PF4 Concentrated Wash recall
PF4 Concentrated Wash (10x), REF 403624, packaged in the following assay kits: 1. ImmucorGamma LIFECODES PF4 IgG (catalog numbers HAT13G, HAT45G) and 2. ImmucorGamma LIFECODES PF4 Enhanced (catalog numbers X-HAT13, X-HAT45) The PF4 IgG assay is an Enzyme Linked Immunosorbent Assay (ELISA). The PF4 IgG ELISA is intended to detect IgG antibodies in human serum that react with Platelet Factor 4 (PF4) when it is complexed to heparin or other polyanionic compounds. The PF4 IgG kit contains all of the reagents necessary to perform the assay.
Why was PF4 Concentrated Wash recalled?
Some Lots of PF4 Concentrated Wash (10X) are a slightly more dilute concentration. This may lead to slightly increased Optical Density (OD) values. Samples with a higher background reactivity may demonstrate false positive result (borderline reactivity) at or near the cutoff for the assay.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Reagent Lot Numbers: 3001620, 3001987, 3002083, 3002153
- Product
- PF4 Concentrated Wash (10x), REF 403624, packaged in the following assay kits: 1. ImmucorGamma LIFECODES PF4 IgG (catalog numbers HAT13G, HAT45G) and 2. ImmucorGamma LIFECODES PF4 Enhanced (catalog numbers X-HAT13, X-HAT45) The PF4 IgG assay is an Enzyme Linked Immunosorbent Assay (ELISA). The PF4 IgG ELISA is intended to detect IgG antibodies in human serum that react with Platelet Factor 4 (PF4) when it is complexed to heparin or other polyanionic compounds. The PF4 IgG kit contains all of the reagents necessary to perform the assay.
- Recalling firm
- Genetic Testing Institute,inc
- Quantity
- 6206 kits
- Distribution
- Worldwide Distribution - Nationwide Distribution to the states of : AL, AZ, AR, CA, CO, CT, DE, DC, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN,TX, UT, VA, WA, WV and WI., and the countries of : Canada, Japan, Belgium, European Union , Australia, Taiwan, Thailand and Scotland. ( some countries not yet identified )
- Recall date
- January 28, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)