FDAClass IIDecember 31, 2025

Beckman Coulter Bicarbonate recall

BECKMAN COULTER BICARBONATE REF: OSR6137 OSR6237 OSR6637 OSR6x37 Bicarbonate is a system reagent for the quantitative determination of Bicarbonate in…

Why was Beckman Coulter Bicarbonate recalled?

Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog Numbers / UDI-DI codes: OSR6137 - 15099590020187 Lot Numbers: 2681 2682 2683 2684 2685 2686 2687 2688 2689 2690 2693 2694 2695 2696 OSR6237 - 15099590010348 Lot Numbers: 2702 2703 2704 2705 2706 2707 2708 2709 2710 2712 2713 2714 2715 2716 OSR6637 - 15099590060206 Lot Numbers: 2726 2728 2729 2730 2731 2732 2733 2734 2736 2737 2738 2739

Product
BECKMAN COULTER BICARBONATE REF: OSR6137 OSR6237 OSR6637 OSR6x37 Bicarbonate is a system reagent for the quantitative determination of Bicarbonate in…
Recalling firm
Beckman Coulter Ireland, Inc.
Quantity
36,981 reagents (U.S.-35,320, O.U.S.-1,661)
Distribution
Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guan, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Argentina, Brazil, Canada, Ecuador, Ireland, Mexico, Netherlands, Panama, Taiwan, and Province of China.
Recall date
December 31, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)