Hemopro 2 with Vasoshield recall: Foreign Material
Hemopro 2 with Vasoshield, Model VH-4001
Why was Hemopro 2 with Vasoshield recalled?
Ink on the Vasoshield syringe Maquet logo may chip resulting in an unreasonable risk of harm to the patient. Potential harms may include foreign body reaction, coronary embolic event, and peripheral embolic event.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI: 00607567700901; Serial Numbers: 25152107, 25152204, 25152293, 25152463, 25152771, 25153153, 25153273, 25153366, 25153585, 25153847, 25153960, 25154066, 25154880, 25155262, 25155326, 25155785, 25156245, 25156642, 25157016, 25157085, 25157244, 25157700, 25157898, 25158066, 25158152, 25158374, 25158822, 25158965, 25159100, 25159335, 25159572, 25159695, 25159791, 25160255, 25160398, 25160498, 25160712, 25160961, 25161193, 25161423, 25161672, 25162202, 25162407, 25162505, 25163168, 25163281, 25163367, 3000236928, 3000251969
- Product
- Hemopro 2 with Vasoshield, Model VH-4001
- Recalling firm
- Maquet Cardiovascular, LLC
- Quantity
- 9817 devices
- Distribution
- US Nationwide. Japan, Germany.
- Recall date
- January 25, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)