BEQ-RF-32-USA RotaFlow Centrifugal Pump with BIOLINE recall: Sterility Concern
BEQ-RF-32-USA RotaFlow Centrifugal Pump with BIOLINE Coating, Product Code 701047554. For use with an extracorporeal cardiovascular or cardiopulmonary…
Why was BEQ-RF-32-USA RotaFlow Centrifugal Pump with BIOLINE recalled?
Nonconformities at their seal could potentially compromise the sterile barrier for affected products. Breach of sterile barrier could expose patients to pathogenic agents, resulting in the following potential harms: inflammation, infection, or sepsis.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI 4037691530864 Lots 3000286570 and 3000325568 ***Lots added 3/21/24*** 3000341070; 3000355577; 3000358977
- Product
- BEQ-RF-32-USA RotaFlow Centrifugal Pump with BIOLINE Coating, Product Code 701047554. For use with an extracorporeal cardiovascular or cardiopulmonary…
- Recalling firm
- Maquet Medical Systems USA
- Quantity
- 1090 total
- Distribution
- Worldwide - US Nationwide distribution to the states of CA, IL, MD, MO, NC, NY, OH, OR, PA, TX, UT, VA and the countries of Canada, Germany, Italy, Japan, and Netherlands.
- Recall date
- February 14, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)