FDAClass IIFebruary 14, 2024

Ingenia Ambition X recall

Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system

Why was Ingenia Ambition X recalled?

Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI 00884838090040 Ingenia Serial number / Accessory Serial Number 47004 100020 47019 100024 47190 100022 47200 100025 47210 100019 47227 100063 47231 100056 47271 100032 47281 100028 47281 100031

Product
Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Recalling firm
Philips North America
Quantity
10
Distribution
Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.
Recall date
February 14, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls