FDAClass IIJanuary 21, 2015

EXp Tibial Insert recall

EXp Tibial Insert ; Intended Use of Device(s): The Proven Gen-Flex Total Knee System is intended for: a. Total knee replacement due to osteoarthritis…

Why was EXp Tibial Insert recalled?

Difficulty locking EXp tibial inserts into tibial trays, caused by a manufacturing condition where a lead in radius on the anterior locking tab is not present on certain insert lots. The absence of this lead in may prevent or make assembly of the insert and tray difficult.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

part numbers SC3425 and SC3453, with lot numbers:28592-103112 28593-102412 28602-110112 28603-110112 28613-110112 28617-110112 28618-110112 28622-112812 28623-112812 28624-112812 31031-121013 31033-121013 31040-121013 31249-022514 31261-020414 28592-103112 28593-102412 28598-111412 28599-111412 28602-110112 28603-110112 28607-111912 28613-110112 28617-110112 28618-110112 28622-112812 28623-112812 28624-112812 31029-120913 31040-121013 31043-120913 31243-022414 31244-022414 31249-022514 31252-021314 31255-020414 31261-020414 32083-080714 31028-120913 31031-121013 31032-121013 31033-121013 31042-121013 31044-120913 31243-022414 31244-022414 31245-022414 31252-021314 31255-020414 31261-020414 31263-021314 31264-031114 31265-031114 31266-031114 28592-103112 28593-102412 28598-111412 28599-111412 28602-110112 28603-110112 28607-111912 28613-110112 28617-110112 28618-110112 28622-112812 28623-112812 28624-112812 31029-120913 31034-120913 31040-121013 31043-120913 31044-120913 31249-022514 31255-020414 32083-080714 28592-103112 28593-102412 28598-111412 28599-111412 28602-110112 28603-110112 28607-111912 28613-110112 28617-110112 28618-110112 28622-112812 28623-112812 28624-112812 31028-120913 31029-120913 31031-121013 31032-121013 31033-121013 31034-120913 31037-121013 31040-121013 31042-121013 31044-120913 31243-022414 31244-022414 31245-022414 31249-022514 31252-021314 31255-020414 31261-020414 31263-021314 31265-031114 32081-080714 28592-103112 28593-102412 28598-111412 28599-111412 28602-110112 28603-110112 28613-110112 28622-112812 28623-112812 28624-112812 31034-120913 28599-111412 28617-110112 31028-120913 31032-121013 31042-121013 31044-120913 31243-022414 31244-022414 31245-022414 31249-022514 31251-022514 31252-021314 31255-020414 31261-020414 31263-021314 31264-031114 31265-031114 31266-031114 28618-110112 28592-103112 28598-103112 28599-111412 28602-110112 28603-110112 28607-111912 28613-110112 28617-110112 28618-110112 28622-112812 28623-112812

Product
EXp Tibial Insert ; Intended Use of Device(s): The Proven Gen-Flex Total Knee System is intended for: a. Total knee replacement due to osteoarthritis…
Recalling firm
Stelkast Co
Quantity
553
Distribution
TX, MD, IN, KS, CA, PA, GA, WA, NY, IL.
Recall date
January 21, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls