FDAClass IIFebruary 19, 2014

BrainLAB's VectorVision spine is intended for use as an recall

BrainLAB's VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It links a…

Why was BrainLAB's VectorVision spine is intended for use as an recalled?

Brainlab has determined that pins of thee Instrument Star Unit ML (Calibration with ICM4) with specific serial numbers might have been made of an incorrect material that is not suitable for the necessary disinfection and sterilization of the product. If corrosion occurs and is not detected by the user, and the device is used during surgery, corroded particles of the pin could, directly or indirect

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model/Catalog Code: 55830-25A Instrument Star Unit ML; Serial Numbers (US Devices): 1205112007, 1205112010, 1205112011, 1205112013, 1205112014, 1205112019, 1205112021, 1205112029, 1205112034, 1205112045; Serial Numbers (Non-US Devices): 1205112004, 1205112006, 1205112008, 1205112009, 1205112015, 1205112016, 1205112017, 1205112018, 1205112020, 1205112023, 1205112025, 1205112026, 1205112027, 1205112030, 1205112032

Product
BrainLAB's VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It links a…
Recalling firm
Brainlab AG
Quantity
25 units
Distribution
Worldwide Distribution - USA including CA, MN, NC, NY, OH, TX and DC. Internationally to Australia, Chile, Germany, Hungary, Italy, Panama, Qatar, Russia, Thailand, United Arab Emirates.
Recall date
February 19, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)