FDAClass IFebruary 8, 2023

CADD Administration Set recall

CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f) REF…

Why was CADD Administration Set recalled?

There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

a) REF 21-7600-24, UDI/DI 30610586044001, Lot Numbers: 4084914 through 4329633; b) REF 21-7609-24, UDI/DI 30610586044018, Lot Numbers: 4072200 through 4334090; c) REF 21-7624-24, UDI/DI 30610586044742, Lot Numbers: 4092506 through 4309481; d) REF 21-7649-24, UDI/DI 35019517184246, Lot Numbers: 4076410 through 4334357; e) REF 21-7600-24JP, UDI/DI N/A, Lot Numbers: 4084914 through 4329633, f) REF 21-7609-24JP, UDI/DI N/A, Lot Numbers: 4072200 through 4334090; g) REF 21-7624-24JP, UDI/DI N/A, Lot Numbers: 4092506 through 4309481; h) REF 21-7649-24JP, UDI/DI N/A, Lot Numbers: 4076410 through 4334357.

Product
CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f) REF…
Recalling firm
Smiths Medical ASD Inc.
Quantity
214668 units
Distribution
Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ, CL, LB, MY, HU, JP, TW, and CN
Recall date
February 8, 2023
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls