FDAClass IIMarch 6, 2019

BRANSIST safire Product Usage: DIGITAL ANGIOGRAPHY recall

BRANSIST safire Product Usage: DIGITAL ANGIOGRAPHY SYSTEM This system is intended to be used for radiologic visualization of the heart, blood vessels…

Why was BRANSIST safire Product Usage: DIGITAL ANGIOGRAPHY recalled?

Normal operation of the device is to power up the device in the morning, register the first patient, and then perform a fluoroscopy. If, however, the user powers up the device in the morning and makes an error by starting a fluoroscopy while the first patient is still being registered, the device application will abnormally terminate and require service intervention before it can be used again. This event will not occur after the first patient procedure.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 40AD58C4B001, 40AD58C55001, 0261Q80503, 0261Q80804, 0161Q84501, 0161Q83102, 3M7A16A1A001, 0161Q82302

Product
BRANSIST safire Product Usage: DIGITAL ANGIOGRAPHY SYSTEM This system is intended to be used for radiologic visualization of the heart, blood vessels…
Recalling firm
Shimadzu Medical Systems Usa Com
Quantity
8 units
Distribution
US Nationwide Distribution in the states of IN, OH, IL, SC
Recall date
March 6, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls