FDAClass IIJanuary 14, 2015

Dimension Vista Myoglobin Calibrator recall

Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic product for the calibration of Myoglobin (MYO) method on…

Why was Dimension Vista Myoglobin Calibrator recalled?

Dimension Vista Myoglobin Calibrator (MYO CAL) may produce a positive shift in MYO QC and Patient test results that exceeds acceptance criteria for this product. A positive shift observed up to 12% at MYO concentrations within and above the reference range of the assay. Depending on quality control limits, this drift may not be detected.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot 4FD085, exp. 12-01-2014

Product
Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic product for the calibration of Myoglobin (MYO) method on…
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
1230
Distribution
Worldwide Distribution: US (nationwide) and country of: Canada.
Recall date
January 14, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)