FDAClass IIDecember 24, 2025

Madsen Accuscreen Teoae/dpoae/abr Probe recall

MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)

Why was Madsen Accuscreen Teoae/dpoae/abr Probe recalled?

Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Part Number: 8-69-41100; UDI-DI: 04260223141355; Serial Numbers: 1008852, 1008888, 1008916, 1008925, 1008935, 1008945, 1008951, 1008957, 1008960, 1008962, 1008966, 1008967, 1008987, 1008990, 1009054, 1009065, 1009066, 1009131, 1009137, 1009138, 1009139, 1009145, 1009148, 1009155, 1009159, 1009161, 1009169, 1009184, 1009188, 1009191, 1009305, 1009315, 1009316, 1009319, 1009321, 1009322, 1009325, 1009328, 1009343, 1009352, 1009356, 1009357, 1009359, 1009366, 1009372, 1009375, 1009376, 1009386, 1009388, 1009390, 1009396, 1009399, 1009400, 1009404, 1009411, 1009415, 1009419, 1009426, 1009427, 1009432, 1009439, 1009442, 1009454, 1009456, 1009460, 1009463, 1009464, 1009468, 1009475, 1009481, 1009488, 1009494, 1009495, 1009497, 1009848, 1009855.

Product
MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)
Recalling firm
PATH
Distribution
Domestic: WI. International: Ireland.
Recall date
December 24, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls