Draeger Infinity Acute Care System recall
Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring…
Why was Draeger Infinity Acute Care System recalled?
The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog Number MS20783 (Infinity MCable-Dual Hemo)
- Product
- Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring…
- Recalling firm
- Draeger Medical, Inc.
- Quantity
- 260
- Distribution
- Distributed in the states of CA, IL, and MA.
- Recall date
- March 20, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)