FDAClass IIIMarch 20, 2013

Siemens Total Iron Binding Capacity recall

Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to quantitatively measure Total Iron Binding…

Why was Siemens Total Iron Binding Capacity recalled?

The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to modify the analytical sensitivity claim to 36 ug/dL (6.44 umol/L).

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Siemens Material Number (SMN) 10444944, Catalog number DF84 -- lot numbers BA3016, EXP 2013-01-16; EC3045, EXP 2013-02-14; EA3073, EXP 2013-03-14; DC3102, EXP 2013-04-12; FC3122, EXP 2013-05-02; EA3157, EXP 2013-06-06; BB3184, EXP 2013-07-03; EC3219, EXP 2013-08-07; EC3220, EXP 2013-08-08; GA3248, 2013-09-05; and FB3290, EXP 2013-10-17

Product
Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to quantitatively measure Total Iron Binding…
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
19, 503 cartons
Distribution
Nationwide Distribution
Recall date
March 20, 2013
Classification
Class III
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls