FDAClass IIFebruary 12, 2014

B. Braun AccuFlo Elastomeric Infusion Pump recall

B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications.

Why was B. Braun AccuFlo Elastomeric Infusion Pump recalled?

Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog #CT-2000-100lots 3H0125F151 and 3H0225F151, expiration 7/2018.

Product
B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications.
Recalling firm
B. Braun Medical, Inc.
Quantity
30, 240 units
Distribution
US Distribution one importer/wholesale distributor in MO.
Recall date
February 12, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls