FDAClass IIFebruary 12, 2014

Crimper recall

Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation.

Why was Crimper recalled?

Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have difficulty fitting a 26mm Sapien transcatheter heart valve into the crimper orifice when preparing the valve for crimping.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

lot 59617777 and 59658495.

Product
Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation.
Recalling firm
Edwards Lifesciences, LLC
Quantity
422
Distribution
Nationwide Distribution
Recall date
February 12, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls