FDAClass IIJanuary 18, 2023

6.5 cm Adult Rotating Craniotome Attachment. Use with recall

6.5 cm Adult Rotating Craniotome Attachment. Use with XMax¿, microMax, and eMax¿ Systems. Ref: CRANI-A Intended for cutting and shaping bone including…

Why was 6.5 cm Adult Rotating Craniotome Attachment. Use with recalled?

Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

GTIN: 00845384001720 All lots

Product
6.5 cm Adult Rotating Craniotome Attachment. Use with XMax¿, microMax, and eMax¿ Systems. Ref: CRANI-A Intended for cutting and shaping bone including…
Recalling firm
The Anspach Effort, Inc.
Quantity
6059 units in total
Distribution
US Nationwide and OUS Foreign: Impacted Foreign Countries Australia Bangladesh China Hongkong India Indonesia Japan Korea (Republic of) Malaysia New Zealand Philippines Singapore Taiwan, Province of China Thailand Angola Austria Belgium Croatia Cyprus Czech Republic Denmark Equatorial Guinea Finland France & French Guiana Georgia Germany Greece Hungary Israel Italy Kenya Kuwait Lebanon Lithuania Macedonia Mauritius Morocco Netherlands Poland Portugal Saudi Arabia Slovakia South Africa Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom of Great Britain and Northern Ireland Argentina Bolivia Brazil Chile Colombia Costa Rica Jamaica Mexico Panama Peru Canada
Recall date
January 18, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)