FDAClass IIFebruary 7, 2024

LARGE PRIMARY REV.TIB.INSERT- DURACON Intended for knee recall

LARGE PRIMARY REV.TIB.INSERT- DURACON Intended for knee replacement Product Number: 6637-4-231

Why was LARGE PRIMARY REV.TIB.INSERT- DURACON Intended for knee recalled?

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

GTIN: 04546540307101 Lot Numbers: 33164901

Product
LARGE PRIMARY REV.TIB.INSERT- DURACON Intended for knee replacement Product Number: 6637-4-231
Recalling firm
Howmedica Osteonics Corp.
Distribution
Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.
Recall date
February 7, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)