FDAClass IDecember 24, 2025

GE Healthcare Carestation anesthesia system recall

GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 750 A1, REF 1012-9750-000; Carestation 750 A2, REF…

Why was GE Healthcare Carestation anesthesia system recalled?

Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

REF 1012-9750-000: GTIN 00840682145596; REF 1012-9750-002: GTIN 00840682146470; REF 1012-9755-000: GTIN 00840682146425; REF 1012-9755-002: GTIN 00840682146463.

Product
GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 750 A1, REF 1012-9750-000; Carestation 750 A2, REF…
Recalling firm
GE Medical Systems China Co., Ltd.
Quantity
888 units
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Argentina, Australia, Bahamas, Bangladesh, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, China, Costa Rica, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lithuania, Malaysia, Mexico, Moldova, Morocco, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Thailand, Tunisia, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Venezuela and Vietnam.
Recall date
December 24, 2025
Classification
Class I
Status
Ongoing
Agency
FDA (device)