FDAClass IIDecember 17, 2025

AMSCO 7053HP Washer/Disinfector recall

AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the…

Why was AMSCO 7053HP Washer/Disinfector recalled?

Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Code: UDI: 00724995176662, 00724995176679, 00724995176655 Serial numbers of distributed units: 3619725008, 3619725011, 3619725013, 3621625010, 3621625011, 3621625013, 3621625014, 3621625015, 3621925001, 3621925002, 3621925005, 3621925010, 3622425005, 3622725005, 3622725011, 3622725012, 3623125006, 3623425016, 3623825008, 3624025002, 3624025008, 3624525007

Product
AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the…
Recalling firm
Steris Corporation
Quantity
22
Distribution
Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia.
Recall date
December 17, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls