Liquid Assayed Specific Protein Control Level 2- In recall
Liquid Assayed Specific Protein Control Level 2- In vitro diagnostic use, in the quality control of serum on clinical chemistry and immunoassay…
Why was Liquid Assayed Specific Protein Control Level 2- In recalled?
Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
GTIN: 05055273204902 Lot Number/Exp. Date: 585LPC 28thJun 23; 601LPC 28th Nov 23; 612LPC 28th Sep 24
- Product
- Liquid Assayed Specific Protein Control Level 2- In vitro diagnostic use, in the quality control of serum on clinical chemistry and immunoassay…
- Recalling firm
- Randox Laboratories Ltd.
- Quantity
- 115 units
- Distribution
- US Nationwide distribution in the states of CA, FL, OH,ME, MO, NC, NJ, PA, TN.
- Recall date
- January 11, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)