Olympus recall
Olympus, Model No. GIF-Q160, flexible scope, large diameter, video gastroscope
Why was Olympus recalled?
If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Numbers: 2407409,2407584,2306291,2204284,2305898,2406901,2306874,2000549,2203971, 2407393,2203790,2000405,2306456,2102668,2203672,2406988,2102128,2103100, 2103035,2204108,2204527,2000464,2204116,2102323,2417865,2102954,2519556, 2418512,2306371,2305350,2407684,2519174,2519207,2519210,2102734,2205217, 2001322,2418634,2406987,2418156,2305446,2204190,2204202,2103307,2101785, 2518969,2519343,2102896,2203946,2001430,2305568,2102316,2305870,2204258
- Product
- Olympus, Model No. GIF-Q160, flexible scope, large diameter, video gastroscope
- Recalling firm
- Integrated Medical Systems Inc
- Quantity
- 54 instruments
- Distribution
- Distributed US (nationwide) and in Australia.
- Recall date
- March 2, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)