FDAClass IIJanuary 14, 2015

ABX PENTRA Reagent Container recall

ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)

Why was ABX PENTRA Reagent Container recalled?

HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. The defect would result in a false high or false low results for HbA1c and false high results for magnes

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL) manufactured since June 2014.

Product
ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)
Recalling firm
Horiba Instruments, Inc dba Horiba Medical
Quantity
443 units
Distribution
Nationwide in US
Recall date
January 14, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)