Philips MX 16-Slice CT Scanner System recall
Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system.
Why was Philips MX 16-Slice CT Scanner System recalled?
The four screws that secure the patient table top to the carrier became loose, which caused the table top to detach.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
System Code #, 728131, Serial #: EP16E090001, EP16E090002, EP16E090003, EP16E090004, EP16E090006, EP16E090007, EP16E090008, EP16E090009, EP16E090010, EP16E090011, EP16E090012,EP16E090013, EP16E090015, EP16E100001, EP16E100004, EP16E100006, EP16E100008, EP16E100010, EP16E100011, EP16E100012, EP16E100013, EP16E100015, EP16E100016, EP16E100017, EP16E100018, EP16E100019, EP16E100020, EP16E100021, EP16E100022, EP16E100023, EP16E100024, EP16E100025, EP16E100026, EP16E100027, EP16E100028, EP16E100029, EP16E100030, EP16E100031, EP16E100032, EP16E100033, EP16E100034, EP16E100035, EP16E100036, EP16E100037, EP16E100038, EP16E100039, EP16E100041, EP16E100043, EP16E100044, EP16E100045, EP16E100046, EP16E100047, EP16E100048, EP16E100049, EP16E100051, EP16E100052, EP16E100055, EP16E100056, EP16E100057, EP16E100058, EP16E100059, EP16E100060, EP16E100061, EP16E100062, EP16E100063, EP16E100067, EP16E100069, EP16E100070, EP16E100071, EP16E100072, EP16E100073, EP16E100074, EP16E100075, EP16E100076, EP16E100077, EP16E100078, EP16E100079, EP16E100080, EP16E100081, EP16E100082, EP16E100083, EP16E100084, EP16E100085, EP16E100086, EP16E100087, EP16E100088, EP16E100090, EP16E100091, EP16E100092, EP16E100093, EP16E100094, EP16E100095, EP16E100096, EP16E100099, EP16E100100, EP16E100101, EP16E100102, EP16E100103, EP16E100105, EP16E100106, EP16E100107, EP16E100108, EP16E100109, EP16E100110, EP16E100111, EP16E100112, EP16E100113, EP16E110003, EP16E110005, EP16E110006, EP16E110007, EP16E110008, EP16E110009, EP16E110011, EP16E110012, EP16E110013, EP16E110014, EP16E110015, EP16E110017, EP16E110018, EP16E110019, EP16E110020, EP16E110021, EP16E110023, EP16E110024, EP16E110025, EP16E110026, EP16E110027, EP16E110028, EP16E110029, EP16E110030, EP16E110031, EP16E110032, EP16E110033, EP16E110034, EP16E110035, EP16E110037, EP16E110038, EP16E110039, EP16E110042, EP16E110043, EP16E110044, EP16E110045, EP16E110046, EP16E110047, EP16E110048, EP16E110049, EP16E110050, EP16E110051, EP16E110052, EP16E110054, EP16E110055, EP16E110056, EP16E110057, EP16E110058, EP16E110060, EP16E110062, EP16E110063, EP16E110064, EP16E110067, EP16E110069, EP16E110070, EP16E110074, EP16E110075, EP16E110077, EP16E110078, EP16E110080, EP16E110083, EP16E110084, EP16E110085, EP16E110089, EP16E110090, EP16E110091, EP16E110092, EP16E110093, EP16E110095, EP16E110096, EP16E110097, EP16E110099, EP16E110100, EP16E110101, EP16E110102, EP16E110104, EP16E110105, EP16E110106, EP16E110107, EP16E110108, EP16E110109, EP16E110110, EP16E110111, EP16E110115, EP16E110117, EP16E110118, EP16E110119, EP16E120003, EP16E120004, EP16E120005, EP16E120006, EP16E120007, EP16E120008, EP16E120009, EP16E120010, EP16E120011, EP16E120013, EP16E120014, EP16E120015, EP16E120016, EP16E120017, EP16E120019, EP16E120021, EP16E120023, EP16E120024, EP16E120025, EP16E120026, EP16E120027, EP16E120029, EP16E120030, EP16E120031, EP16E120032, EP16E120033, EP16E120034, EP16E120035, EP16E120036, EP16E120037, EP16E120039, EP16E120040, EP16E120041, EP16E120042, EP16E120043, EP16E120044, EP16E120045, EP16E120046, EP16E120047, EP16E120048, EP16E120050, EP16E120051, EP16E120052, EP16E120055, EP16E120056, EP16E120058, EP16E120059, EP16E120060, EP16E120061, EP16E120063, EP16E120065, EP16E120067, EP16E120068, EP16E120070, EP16E120071, EP16E120072, EP16E120074, EP16E120075, EP16E120076, EP16E120078, EP16E120081, EP16E120083, EP16E120088, EP16E120089, EP16EB110003, EP16EB110004, EP16EB110005, EP16EB110006, P16C080001, P16C080002, P16C100001, P16C100002, P16C100003, P16C100004, P16C100005, P16C100007, P16C100008, P16C100009, P16C100010, P16C100011, P16C100012, P16C100013, P16C100014, P16C100015, P16C100017, P16C100018, P16C110001, P16C110002, P16C110003, P16C110004, P16C110005, P16C110006, P16C110007, P16C110008, P16C110011, P16C110012, P16C110013, P16C110014, P16C110015, P16C110016, P16C110017, P16C110018, P16C110019, P16C110020, P16C110021, P16C110022, P16C110023, P16C110024, P16C110025, P16C110026, P16C110027, P16C110028, P16C110029, P16C110030, P16C110031, P16C110032, P16C110033, P16C110034, P16C110035, P16C110036, P16C110037, P16C120001, P16C120002, P16C120003, P16C120006, P16C120007, P16C120008, P16C120009, P16C120011, P16C120012, P16C120014, P16C120016, P16C120017, P16C120018, P16C120019, P16C120020, P16C120021, P16C120022, P16C120023, P16C120024, P16C120025, P16C120026, P16C120027, P16C120029, P16C120030, P16E080007, P16E080008, P16E090002, P16E090003, P16E090004, P16E090005, P16E090006, P16E090008, P16E090009, P16E090010, P16E090011, P16E090012, P16E090013, P16E090014, P16E090015, P16E090016, P16E090017, P16E090018, P16E090019, P16E090020, P16E090021, P16E090024, P16E090025, P16E090026, P16E090027, P16E090028, P16EF80010, P16EM08003, P16EM09002, P16EM09003, P16EM09004 System Code #, 728132, Serial #: EP16E100005, EP16E100007, EP16E110066, EP16ER100003, EP16ER100004, EP16ER100005, EP16ER100007, EP16ER100008, EP16ER110003, EP16ER110004, EP16ER110005, EP16ER110006, EP16ER110007, EP16ER110008, EP16ER110009, EP16ER110010, EP16ER110011, EP16ER110012, EP16ER110013, EP16ER110014, EP16ER110015, EP16ER110016, EP16ER110017, EP16ER110018, EP16ER110019, EP16ER110020, EP16ER110021, EP16ER110022, EP16ER110023, EP16ER120005, EP16ER120006, EP16ER120007, EP16ER120008, EP16ER120009, EP16ER120010, EP16ER120011, EP16ER120012, EP16ER120013, EP16ER120015, EP16ER120016 , EP16ER120017, EP16ER120018
- Product
- Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system.
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Quantity
- 412 Units
- Distribution
- Worldwide Distribution - USA including AL, AR, CA, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NV, NY, OH, PA, PR and VA.
- Recall date
- March 6, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)