FDAClass IIMarch 14, 2018

Auriga XL 4007 Laser System recall

Auriga XL 4007 Laser System, Universal Part Number: M0068FS4007G0. Intended to be used in surgical procedures.

Why was Auriga XL 4007 Laser System recalled?

Certain Auriga consoles have incorrectly assembled trigger wire components. The trigger wire functions in the start-up of the console. An incorrectly assembled trigger wire may result in error codes or failure to start the console.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Number: 20509715, 20465096, 20503137, 20520960, 20541967, 20546219, 20546440, 20575512, and 20575518.

Product
Auriga XL 4007 Laser System, Universal Part Number: M0068FS4007G0. Intended to be used in surgical procedures.
Recalling firm
Boston Scientific Corporation
Quantity
9
Distribution
International Distribution Only -- Australia, Vietnam, France, Germany, Italy, Spain, and Romania.
Recall date
March 14, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls